Development of a MR compatible brain PET II using 4-side tileable GAPD arrays
Author :
Jin Ho Jung ; Yong Choi ; Jiwoong Jung ; Sangsu Kim ; Hyun Keong Lim ; Ki Chun Im ; Hyeok-jun Choe ; Yoonsuk Huh ; Kyu Born Kim ; Chang Hyun Oh ; Kyung Min Kim ; Jong Guk Kim ; Hyun-Wook Park
Author_Institution :
Dept. of Electron. Eng., Sogang Univ., Seoul, South Korea
fYear :
2012
fDate :
Oct. 27 2012-Nov. 3 2012
Firstpage :
2783
Lastpage :
2787
Abstract :
There has been great interest in the development of combined PET/MR, a useful tool for both functional and anatomic imaging. We have developed a proof-of-principle MR compatible PET, employing a design concept that uses GAPD arrays as a PET photo-sensor and charge signal transmission method for human brain imaging. The purpose of this study was to design the 2nd version of brain PET with an extended axial field-of-view (FOV) and to evaluate its initial performance. The PET consisted of 18 detector blocks arranged in a ring of 390 mm diameter with 60 mm axial FOV. Each detector block was composed of a 4 × 4 matrix of detector module, each of which consisted of a 4 × 4 array L YSO coupled to a 4-side tileable 4 × 4 GAPD array. The PET gantry was shielded with gold-plated conductive fabric tapes with a thickness of 0.1 mm. PET signals were fed into the position decoder circuit (PDC) generating the digital address and analog pulse of the one interacted channel among the 256 output channels of the detector block, using a 4 m long flat cable. Commercial DAQ modules were used to digitize analog output signals of the PDCs and to store the data in list mode format. The flat cable was shielded with a mesh-type aluminum sheet, which had a thickness of 0.24 mm. All electronics were enclosed in an aluminum box, which had a thickness of 10 mm, located outside the MR bore. Average energy and timing resolutions of the developed PET measured outside the MR room were 18.1±3.2% (n=4,608) and 3.6 ns, respectively. The sensitivity and spatial resolution were 1.2% and 3.1 mm at the center of the field of view, respectively. No significant degradations of PET performance and the uniformity of MR image were observed. Simultaneous PET and MR images of hot-rod phantom and cat brain were successfully acquired. Experimental results indicate that the high performance compact and lightweight PET insert for hybrid PET-MRI can be developed using GAPD arrays and cha- ge signal transmission method.
Keywords :
biomedical MRI; brain; image sequences; medical image processing; phantoms; positron emission tomography; GAPD arrays; MR compatible brain PET II; MR image sequence; PET gantry; PET photosensor; aluminum box; analog pulse; anatomic imaging; cat brain; charge signal transmission method; commercial DAQ modules; digital address; extended axial field-of-view; four-side tileable GAPD arrays; functional imaging; gold-plated conductive fabric tapes; hot-rod phantom; human brain imaging; mesh-type aluminum sheet; position decoder circuit generation; proof-of-principle MR compatible PET; size 0.1 mm; size 0.24 mm; size 10 mm; size 390 mm; size 4 m; size 60 mm; time 3.6 ns;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC), 2012 IEEE
Conference_Location :
Anaheim, CA
ISSN :
1082-3654
Print_ISBN :
978-1-4673-2028-3
Type :
conf
DOI :
10.1109/NSSMIC.2012.6551635
Filename :
6551635
Link To Document :
بازگشت