DocumentCode :
2558783
Title :
Sensitivity improvement of time-of-flight (ToF)-PET detector through recovery of Compton scattered annihilation photons
Author :
Wagadarikar, Ashwin A. ; Ivan, Ash ; Dolinsky, Sergei ; McDaniel, D.L.
Author_Institution :
GE Global Res., Niskayuna, NY, USA
fYear :
2012
fDate :
Oct. 27 2012-Nov. 3 2012
Firstpage :
3178
Lastpage :
3183
Abstract :
In PET detector designs, the scintillator material can be partitioned such that a block of crystals share timing/energy readout electronics. A fraction of the incident 511 keY photons produce simultaneous events in two adjacent block readouts due to Compton scattering followed by escape from the primary block. These inter-block Compton scatter events are typically not processed in current PET scanners. We have used radiation transport simulations to determine the fraction of inter-block Compton events for different block sizes using LYSO scintillator. The simulations showed the statistical distribution of the energy signals and time stamps in the two blocks and guided the selection of energy and time criteria for an event recovery algorithm. With a particular block size, we experimentally demonstrated that inter-block Compton events may be recovered as valid events with a corrected time stamp and estimated position of initial interaction. The results showed a significant improvement in detector sensitivity at the expense of a small degradation in the timing resolution of the detector block.
Keywords :
Compton effect; nuclear electronics; positron emission tomography; readout electronics; solid scintillation detectors; Compton scattered annihilation photons; LYSO scintillator; PET scanners; block readouts; block sizes; corrected time stamp; detector sensitivity improvement; energy readout electronics; energy selection; energy signal statistical distribution; event recovery algorithm; initial interaction position; inter-block Compton scatter events; radiation transport simulations; time-of-flight-PET detector; timing readout electronics;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC), 2012 IEEE
Conference_Location :
Anaheim, CA
ISSN :
1082-3654
Print_ISBN :
978-1-4673-2028-3
Type :
conf
DOI :
10.1109/NSSMIC.2012.6551725
Filename :
6551725
Link To Document :