Calibration methodology for a dual-ended readout silicon photomultiplier based depth-of-interaction PET detector module
Author
Xi, Wei ; McKisson, J.E. ; Weisenberger, Andrew G. ; Lee, Sang-Rim ; Taylor, Mark ; Stepanyan, A. ; Zorn, Chris
Author_Institution
Jefferson Lab., Newport News, VA, USA
fYear
2012
fDate
Oct. 27 2012-Nov. 3 2012
Firstpage
3493
Lastpage
3495
Abstract
We developed a novel calibration methodology for a PET detector with dual-ended readout of an LYSO array by two silicon photomultipliers (SiPMs). By introducing a detector gain balancing step in the calibration process, improved depth-of-interaction calibration uniformity and accuracy can be achieved. The entire calibration process has four steps: scintillation crystal array mappings for two SiPM readouts, detector gain balancing, energy calibration, and depth-of-interaction calibration. This document provides a detailed description on the detector calibration system setup.
Keywords
calibration; photomultipliers; positron emission tomography; readout electronics; silicon radiation detectors; LYSO array; PET detector module; accuracy; calibration methodology; depth-of-interaction calibration; detector gain balancing step; dual ended readout; energy calibration; scintillation crystal array mappings; silicon photomultiplier; uniformity;