    Calibration methodology for a dual-ended readout silicon photomultiplier based depth-of-interaction PET detector module

  • Author

    Xi, Wei ; McKisson, J.E. ; Weisenberger, Andrew G. ; Lee, Sang-Rim ; Taylor, Mark ; Stepanyan, A. ; Zorn, Chris

  • Author_Institution
    Jefferson Lab., Newport News, VA, USA
  • fYear
    2012
  • fDate
    Oct. 27 2012-Nov. 3 2012
  • Firstpage
    3493
  • Lastpage
    3495
  • Abstract
    We developed a novel calibration methodology for a PET detector with dual-ended readout of an LYSO array by two silicon photomultipliers (SiPMs). By introducing a detector gain balancing step in the calibration process, improved depth-of-interaction calibration uniformity and accuracy can be achieved. The entire calibration process has four steps: scintillation crystal array mappings for two SiPM readouts, detector gain balancing, energy calibration, and depth-of-interaction calibration. This document provides a detailed description on the detector calibration system setup.
  • Keywords
    calibration; photomultipliers; positron emission tomography; readout electronics; silicon radiation detectors; LYSO array; PET detector module; accuracy; calibration methodology; depth-of-interaction calibration; detector gain balancing step; dual ended readout; energy calibration; scintillation crystal array mappings; silicon photomultiplier; uniformity;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC), 2012 IEEE
  • Conference_Location
    Anaheim, CA
  • ISSN
    1082-3654
  • Print_ISBN
    978-1-4673-2028-3
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2012.6551797
  • Filename
    6551797