DocumentCode
2560424
Title
Development of a new brain PET scanner based on single event data acquisition
Author
Isobe, Takanori ; Yamada, Ryota ; Shimizu, Kazuo ; Saito, Akihiro ; Ote, K. ; Sakai, Kenji ; Moriya, Takehiro ; Yamauchi, Hiroyuki ; Omura, T. ; Watanabe, Manabu
Author_Institution
Central Reserch Lab., Dev. Center of Hamamatsu Photonics K.K., Hamamatsu, Japan
fYear
2012
fDate
Oct. 27 2012-Nov. 3 2012
Firstpage
3540
Lastpage
3543
Abstract
A new brain PET scanner based on single event data acquisition with on-the-fly coincidence detection has been developed. The new scanner was designed to form a detector ring of 430 mm in diameter with 32 detector modules. The single event data generated at each detector module were transferred to the data acquisition system through a fiber cable. The single event data from all detector modules were merged and processed to create coincidence event data in software with on-the-fly. The validation results indicated that the single count rate capability was expected over 220 Mcps. At experiments using 18F radioisotopes, the maximum single count rate was approximately 20 Mcps, and the maximum coincidence count rate with prompt was obtained 3.2 Mcps. The system timing resolution was 2.6 ns FWHM, and the optimum coincidence time window was 4.3 ns, obtained by adjusting the timing delay variations between each detector module.
Keywords
data acquisition; image reconstruction; medical image processing; positron emission tomography; radioisotopes; brain PET scanner; detector module; detector modules; image reconstruction; on-the-fly coincidence detection; radioisotopes; single event data acquisition system; size 430 mm; timing delay variations;
fLanguage
English
Publisher
ieee
Conference_Titel
Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC), 2012 IEEE
Conference_Location
Anaheim, CA
ISSN
1082-3654
Print_ISBN
978-1-4673-2028-3
Type
conf
DOI
10.1109/NSSMIC.2012.6551810
Filename
6551810
Link To Document