DocumentCode
2560589
Title
Ultrasonic parameters of renal calculi
Author
Singh, V.R. ; Singh, Suresh
Author_Institution
Nat. Phys. Lab., New Delhi, India
Volume
2
fYear
1998
fDate
29 Oct-1 Nov 1998
Firstpage
862
Abstract
Ultrasonic parameters such as ultrasonic velocity, acoustic impedance, attenuation, dynamic modulus of elasticity, coefficient of reflectance and transmittance have been determined for several renal calculi samples, in vitro. These parameters are studied by using a single probe method (pulse echo) in liquid form at different concentrations (gm/cc). It was found that as the concentration increases, the ultrasonic velocity increases. A pulse receiver (model Panametries-5052 PR) was used to excite the transducer (2.5 MHz) and the output was given to CRO (model-OS-300C, 20 MHz). A sample holder assembly was used in the present work. The authors have also determined all acoustic parameter mentioned above by using a double-probe through-transmission method for several renal calculi samples in the solid rock form. The present study would help in determining the design parameters to design and develop a simple kidney stone lithotripter
Keywords
acoustic impedance; biomedical ultrasonics; elastic moduli; kidney; ultrasonic reflection; ultrasonic transmission; ultrasonic velocity; 20 MHz; design parameters determination; elasticity dynamic modulus; model Panametries-5052 PR; model-OS-300C; pulse receiver; renal calculi; simple kidney stone lithotripter; solid rock form; ultrasonic parameters; Acoustic pulses; Acoustic transducers; Attenuation; Elasticity; Impedance; In vitro; Kidney stones; Probes; Reflectivity; Ultrasonic transducers;
fLanguage
English
Publisher
ieee
Conference_Titel
Engineering in Medicine and Biology Society, 1998. Proceedings of the 20th Annual International Conference of the IEEE
Conference_Location
Hong Kong
ISSN
1094-687X
Print_ISBN
0-7803-5164-9
Type
conf
DOI
10.1109/IEMBS.1998.745574
Filename
745574
Link To Document