    AC-DC voltage transfer difference of the PTB-multijunction thermal converter at 1 MHz

  • Author

    Klonz, M.

  • Author_Institution
    Phys.-Tech. Bundesanstalt, Braunschweig, Germany
  • fYear
    1990
  • fDate
    11-14 June 1990
  • Firstpage
    60
  • Lastpage
    61
  • Abstract
    The AC-DC voltage transfer difference of a 700- Omega -heater multijunction thermal converter has been evaluated in the frequency range from 100 kHz to 1 MHz using the same schematic diagram developed for frequencies below 100 kHz. The agreement between calculated and measured values with increasing transfer differences for lower heater resistances indicates the validity of the schematic diagram and the estimation procedure. The estimated uncertainty at 1 MHz is 10 p.p.m.<>
  • Keywords
    convertors; voltage measurement; 100 kHz to 1 MHz; 700 ohm; AC-DC voltage transfer; PTB-multijunction thermal converter; international comparison; Analog-digital conversion; Connectors; Dielectric losses; Frequency conversion; Frequency estimation; Heat transfer; Standards publication; Uncertainty; Voltage; Wire;