• DocumentCode
    2561169
  • Title

    Recent activities at BIPM in the field of stabilized lasers: radiations recommended for the definition of the metre

  • Author

    Chartier, J.-M. ; Robertsson, L. ; Fredin-Picard, S.

  • Author_Institution
    Bur. Int. des Poids et Mesures, Sevres, France
  • fYear
    1990
  • fDate
    11-14 June 1990
  • Firstpage
    94
  • Lastpage
    95
  • Abstract
    The results of interconnectional comparisons of iodine-stabilized He-Ne lasers at lambda =633 nm between Bureau International des Poids et Mesures (BIPM) and 11 national laboratories are presented. The measured frequency differences were all within +or-3 parts in 10/sup 11/. Investigations concerning He-Ne lasers at lambda =612 nm using internal or external iodine cells are discussed. The frequency reproducibility is limited by, among other things the quality of the iodine cells. Therefore, the contamination of iodine cells has been studied using laser-induced fluorescence. First results obtained on the green line of He-Ne at lambda =543 nm stabilized on iodine are presented. The frequency repeatability between two such systems was (14+or-5) kHz.<>
  • Keywords
    gas lasers; helium; iodine; laser beam applications; laser frequency stability; length measurement; neon; units (measurement); 543 nm; 612 nm; 633 nm; BIPM; He-Ne lasers; I/sub 2/ stabilised laser; contamination; external iodine cells; frequency repeatability; frequency reproducibility; green line; internal iodine cell; laser-induced fluorescence; metre; stabilized lasers; Absorption; Contamination; Extraterrestrial measurements; Fluorescence; Frequency measurement; Laboratories; Laser stability; Laser theory; Metrology; Pollution measurement;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Precision Electromagnetic Measurements, 1990. CPEM '90 Digest., Conference on
  • Conference_Location
    Ottawa, Ontario, Canada
  • Type

    conf

  • DOI
    10.1109/CPEM.1990.109939
  • Filename
    109939