• DocumentCode
    2561527
  • Title

    SiPM-based PET module with depth of interaction

  • Author

    Pennazio, F. ; Ambrosi, G. ; Bisogni, Maria G. ; Cerello, P. ; Corsi, F. ; Del Guerra, Alberto ; Ionica, Mircea Horea ; Marino, Nahema ; Marzocca, Cristoforo ; Morrocchi, Matteo ; Peroni, C. ; Pirrone, G. ; Santoni, C. ; Wheadon, Richard

  • Author_Institution
    Dept. of Phys., Univ. of Torino, Turin, Italy
  • fYear
    2012
  • fDate
    Oct. 27 2012-Nov. 3 2012
  • Firstpage
    3786
  • Lastpage
    3789
  • Abstract
    Silicon Photomultipliers are used in many new generation PET block detectors. High granularity pixel SiPMs allow a high precision measurement of the photon interaction coordinates along the crystal surface. In order to further improve the resolution it is necessary to measure the photon Depth of Interaction (DOl), so as to reduce the parallax error in the Line of Response reconstruction. An innovative technique for DOl determination is proposed and tested. Measurements are made with a 2 cm × 2 cm × 1 cm LYSO slab with readout on the front and back large sides by means of two 4 × 4 square SiPM pixel matrices of 5 mm pitch. The data acquisition is based on the new BASIC32 chip read out with an FPGA-based system.
  • Keywords
    data acquisition; elemental semiconductors; modules; photomultipliers; positron emission tomography; silicon; silicon radiation detectors; BASIC32 chip read out; DOl; FPGA-based system; LYSO slab; PET block detectors; PET module; Si; SiPM; data acquisition; depth-of-interaction; line-of-response reconstruction; silicon photomultipliers; size 1 cm; size 2 cm;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC), 2012 IEEE
  • Conference_Location
    Anaheim, CA
  • ISSN
    1082-3654
  • Print_ISBN
    978-1-4673-2028-3
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2012.6551869
  • Filename
    6551869