Abstract
A microcalorimeter for dynamic measurement based on the adaptive control of RF-DC substitution is described. The method has allowed a significant simplification of the calorimeter´s hardware and a remarkable speedup of the measurement. The experiments led to a rate of up to 20 with an acceptable uncertainty increase; in addition, the demand for the suppression of perturbations was reduced.<>
Keywords
calorimeters; RF-DC substitution; adaptive control; calorimeter; dynamic measurement; microcalorimeter; nonsteady operation; suppression of perturbations; Adaptive control; Bolometers; Circuits; Copper; Filters; Microwave measurements; Power measurement; Temperature measurement; Time measurement; Voltage;