DocumentCode :
2566395
Title :
A 32-Site 4-Channel Cochlear Electrode Array
Author :
Bhatti, Pamela T. ; Lee, Sangwoo ; Wise, Kensall D.
Author_Institution :
Michigan Univ., Ann Arbor, MI
fYear :
2006
fDate :
6-9 Feb. 2006
Firstpage :
79
Lastpage :
88
Abstract :
A thin-film cochlear electrode array has been developed to improve pitch perception and reduce insertion damage. A silicon-parylene substrate supports 32 IrO sites on 250mum centers along with circuitry for current generation and position sensing. Interfacing over eight leads at plusmn3V, stimulus pulses cover plusmn500muA with 8b resolution and a minimum pulse width of 4mus
Keywords :
biomedical electrodes; biosensors; ear; iridium compounds; silicon compounds; 250 micron; IrO; cochlear electrode array; current generation; insertion damage; pitch perception; position sensing; silicon parylene substrate; Adaptive arrays; Animals; Circuits; Dielectric substrates; Electrodes; Etching; Humans; Sensor arrays; Silicon; Wire;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Solid-State Circuits Conference, 2006. ISSCC 2006. Digest of Technical Papers. IEEE International
Conference_Location :
San Francisco, CA
ISSN :
0193-6530
Print_ISBN :
1-4244-0079-1
Type :
conf
DOI :
10.1109/ISSCC.2006.1696036
Filename :
1696036
Link To Document :
بازگشت