DocumentCode :
2566465
Title :
A 60/spl mu/W 60 nV/Hz Readout Front-End for Portable Biopotential Acquisition Systems
Author :
Yazicioglu, Refet Firat ; Merken, P. ; Puers, Robert ; Van Hoof, Chris
Author_Institution :
IMEC, Leuven
fYear :
2006
fDate :
6-9 Feb. 2006
Firstpage :
109
Lastpage :
118
Abstract :
A biopotential readout front-end can be configured to extract EEG, ECG, and EMG signals and draws 20muA from 3V. AC coupling of chopped amplifiers results in an input-referred noise of 60nV/radicHz and CMRR of 120dB at 1kHz. The immunity of the CMRR to electrode offset voltages is improved with an active input stage and 110dB CMRR is achieved at 100Hz with 50mV electrode offset
Keywords :
biomedical electrodes; data acquisition; electrocardiography; electroencephalography; electromyography; medical signal processing; portable instruments; readout electronics; 1 kHz; 100 Hz; 50 mV; 60 muW; CMRR; ECG; EEG; EMG; electrode offset voltages; portable biopotential acquisition systems; readout front-end; Choppers; Electrocardiography; Electrodes; Electroencephalography; Electromyography; Frequency; Immune system; Interference; Power dissipation; Voltage;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Solid-State Circuits Conference, 2006. ISSCC 2006. Digest of Technical Papers. IEEE International
Conference_Location :
San Francisco, CA
ISSN :
0193-6530
Print_ISBN :
1-4244-0079-1
Type :
conf
DOI :
10.1109/ISSCC.2006.1696039
Filename :
1696039
Link To Document :
بازگشت