DocumentCode
2572086
Title
Calibration of C-arm for orthopedic interventions via statistical model-based distortion correction and robust phantom detection
Author
Schumann, Steffen ; Dong, Xiao ; Puls, Marc ; Nolte, Lutz-Peter ; Zheng, Guoyan
Author_Institution
Inst. for Surg. Technol. & Biomech., Univ. of Bern, Bern, Switzerland
fYear
2012
fDate
2-5 May 2012
Firstpage
1204
Lastpage
1207
Abstract
In this paper, we present a novel C-arm calibration approach for orthopedic applications that is based on a distortion correction plate with fiducials arranged in a single-plane and a novel mobile phantom. The former is designed to be rigidly attached to the image intensifier and is used together with a statistical model of the geometrical image distortion pattern for intra-operative image distortion correction. The latter can be placed to anywhere close to the patient anatomy being imaged. A simulation based method is then proposed for a robust detection of the phantom from an intra-operatively acquired image in order to compute both the intrinsic and the extrinsic parameters of the acquired image. Validation study performed on 20 cadaver images demonstrated an average calibration accuracy of 0.8 ± 0.4 mm.
Keywords
calibration; diagnostic radiography; medical image processing; orthopaedics; phantoms; statistical analysis; C-arm calibration; distortion correction plate; fiducials; geometrical image distortion pattern; image intensifier; intraoperative image distortion correction; mobile phantom; orthopedic interventions; robust phantom detection; statistical model based distortion correction; Accuracy; Buildings; Calibration; Computational modeling; Mobile communication; Phantoms; Vectors; X-rays; calibration; orthopedic procedures;
fLanguage
English
Publisher
ieee
Conference_Titel
Biomedical Imaging (ISBI), 2012 9th IEEE International Symposium on
Conference_Location
Barcelona
ISSN
1945-7928
Print_ISBN
978-1-4577-1857-1
Type
conf
DOI
10.1109/ISBI.2012.6235777
Filename
6235777
Link To Document