A simple set up for the measurement of patient response time in fMRI
Author :
Demir, Ali ; Keskin, Ali Ümit ; Ünsal, Mehmet ; Bayram, Ali
Author_Institution :
Radyoloji Ana Bilim Dali, Yeditepe Univ. Hastanesi, İstanbul, Turkey
Abstract :
Because blood oxygen level dependent (BOLD) functional magnetic resonance imaging (fMRI) pulse sequences have rapidly changing high field gradients, almost all currently available equipment which measure and record the response of a patient to visual or audio stimuli in fMRI employ fiber-optics based systems. In this work we introduce a simple and low cost patient response set up which does not require fiber-optics based sensors and interface. The basis of the method is to transmit an on/off light signal through the RF shielded window of the scanning room. The performance of the proposed measurement set up is tested using fMRI sequences in a 3 Tesla MRI unit.
Keywords :
biomedical MRI; biomedical equipment; fibre optic sensors; optoelectronic devices; BOLD functional magnetic resonance imaging; RF shielded window; audio stimuli; blood oxygen level dependent fMRI; fMRI pulse sequence; fiber optic interface; fiber optic sensors; fiber-optics based systems; patient response time; visual stimuli; Blood; Costs; Current measurement; Delay; Magnetic field measurement; Magnetic resonance imaging; Optical fiber devices; Optical fiber sensors; Pulse measurements; Time measurement;