DocumentCode
2592119
Title
Vibrating wire sensor measurement method by stimulation with steps of variable frequency sinusoidal pulse trains
Author
Istvan, Kovacs ; Maria, Mircea ; Bela-Zoltan, Gyurka ; Szabolcs, Balogh
Author_Institution
S.C. IPA S.A., Cluj-Napoca, Romania
fYear
2012
fDate
24-27 May 2012
Firstpage
587
Lastpage
590
Abstract
In some cases, “pluck-read” (single coil) measurement mode of the resonant frequency of vibrating wire sensors does not lead to result. For these cases, the authors of this article have developed a new stimulation method with step of variable frequency sinusoidal pulse trains. As a result, the amortized output signal duration of the sensor will be longer, allowing the measurement of some, partially deteriorated vibrating wire sensors, which cannot be measured with other methods.
Keywords
sensors; vibration measurement; amortized output signal duration; pluck-read measurement mode; single coil measurement mode; variable frequency sinusoidal pulse trains; vibrating wire sensor measurement method; Coils; Resonant frequency; Time frequency analysis; Time measurement; Vibration measurement; Wires; resonant frequency; sensor; vibrating wire;
fLanguage
English
Publisher
ieee
Conference_Titel
Automation Quality and Testing Robotics (AQTR), 2012 IEEE International Conference on
Conference_Location
Cluj-Napoca
Print_ISBN
978-1-4673-0701-7
Type
conf
DOI
10.1109/AQTR.2012.6237778
Filename
6237778
Link To Document