• DocumentCode
    2592119
  • Title

    Vibrating wire sensor measurement method by stimulation with steps of variable frequency sinusoidal pulse trains

  • Author

    Istvan, Kovacs ; Maria, Mircea ; Bela-Zoltan, Gyurka ; Szabolcs, Balogh

  • Author_Institution
    S.C. IPA S.A., Cluj-Napoca, Romania
  • fYear
    2012
  • fDate
    24-27 May 2012
  • Firstpage
    587
  • Lastpage
    590
  • Abstract
    In some cases, “pluck-read” (single coil) measurement mode of the resonant frequency of vibrating wire sensors does not lead to result. For these cases, the authors of this article have developed a new stimulation method with step of variable frequency sinusoidal pulse trains. As a result, the amortized output signal duration of the sensor will be longer, allowing the measurement of some, partially deteriorated vibrating wire sensors, which cannot be measured with other methods.
  • Keywords
    sensors; vibration measurement; amortized output signal duration; pluck-read measurement mode; single coil measurement mode; variable frequency sinusoidal pulse trains; vibrating wire sensor measurement method; Coils; Resonant frequency; Time frequency analysis; Time measurement; Vibration measurement; Wires; resonant frequency; sensor; vibrating wire;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Automation Quality and Testing Robotics (AQTR), 2012 IEEE International Conference on
  • Conference_Location
    Cluj-Napoca
  • Print_ISBN
    978-1-4673-0701-7
  • Type

    conf

  • DOI
    10.1109/AQTR.2012.6237778
  • Filename
    6237778