Title :
Researches on a new dry calibration method of electromagnetic flowmeter based on unit element
Author :
Sun Xiangdong ; Libin ; Zhao zifeng
Author_Institution :
Autom. Coll., Shanghai Univ., Shanghai, China
Abstract :
The need for calibration of large electromagnetic flowmeters(EMF) adds considerably to the cost of these instruments. In order to reduce the cost, this paper discusses a technique for calibration of large electromagnetic flowmeters. The proposed technique based on unit element, moving unit flowrate set to measure the electromotive force of every point in the tube, and adds the weighting function arithmetic to get a complete set of integrity theory. Now a set of dry calibration equipment have been produced, and using this equipment accomplished the calibration of DN300 electromagnetic flowmeter, the calibration results accord with former expectation.
Keywords :
calibration; electromagnetic devices; flowmeters; pipe flow; DN300 EMF; dry calibration method; electromagnetic flowmeter; electromotive force measurement; integrity theory; pipeline flowrate; unit element; weighting function arithmetic; Calibration; Costs; Current; Electrodes; Electromagnetic forces; Electromagnetic measurements; Fluid flow measurement; Force measurement; Magnetic field measurement; Space technology; Electromagnetic Flowmeter; Unit element; dry calibration;
Conference_Titel :
Instrumentation and Measurement Technology Conference, 2009. I2MTC '09. IEEE
Conference_Location :
Singapore
Print_ISBN :
978-1-4244-3352-0
DOI :
10.1109/IMTC.2009.5168457