• DocumentCode
    2599519
  • Title

    Harmonization of standard procedures for ADC time domain dynamic testing

  • Author

    Balestrieri, E. ; Daponte, P. ; Rapuano, Sergio

  • Author_Institution
    Dept. of Eng., Univ. of Sannio, Benevento, Italy
  • fYear
    2009
  • fDate
    5-7 May 2009
  • Firstpage
    1238
  • Lastpage
    1243
  • Abstract
    The paper deals with the ADC dynamic parameters and test methods in the time domain by means of sine wave fitting. An analytical comparison of these methods, according to the internationally available standards and the research in the field, is described. The results of an experimental investigation carried out on actual ADCs by applying the sine wave fitting methods are also reported and compared as a step towards the harmonization of the existing standards.
  • Keywords
    analogue-digital conversion; dynamic testing; time-domain analysis; ADC time domain dynamic testing; sine wave fitting; standard procedure harmonization; Algorithm design and analysis; Fitting; IEC standards; Instrumentation and measurement; Manufacturing; Measurement standards; Paper technology; Phase measurement; Testing; Time measurement; ADC; DYNAD; ENOB; IEC; IEEE; SINAD; SNR; THD; harmonization; sine-fit;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Instrumentation and Measurement Technology Conference, 2009. I2MTC '09. IEEE
  • Conference_Location
    Singapore
  • ISSN
    1091-5281
  • Print_ISBN
    978-1-4244-3352-0
  • Type

    conf

  • DOI
    10.1109/IMTC.2009.5168644
  • Filename
    5168644