• DocumentCode
    260058
  • Title

    Electromagnetic actuation system for locomotive intravascular therapeutic microrobot

  • Author

    Hyunchul Choi ; Gwangjun Go ; Cheong Lee ; Seong Young Ko ; Semi Jeong ; Kiduk Kwon ; Jong-Oh Park ; Sukho Park

  • Author_Institution
    Sch. of Mech. Eng., Chonnam Nat. Univ., Gwangju, South Korea
  • fYear
    2014
  • fDate
    12-15 Aug. 2014
  • Firstpage
    831
  • Lastpage
    834
  • Abstract
    In this paper, we proposed an intravascular therapeutic microrobot using an electromagnetic actuation (EMA) system with bi-plane X-ray imaging device. The proposed EMA system consists of Helmholtz-Maxwell coils, uniform-gradient saddle coils. The Helmholtz-Maxwell coils are located along y-axis, and uniform-gradient saddle coils are located perpendicular to y-axis. In order to align the microrobot along a desired angle in 2D (dimensional) plane, it is necessary to control of the currents on Helmholtz coil and uniform saddle coil. For a forward and backward direction movement of the microrobot, we precisely control the currents of Maxwell coil and gradient saddle coil. Because the saddle coils can be rotated around the y-axis, the effective actuation plane of the microrobot can be also rotated, and the microrobot can move in 3D space. In addition, for the position recognition of the microrobot in a blood vessel, we adopted a bi-plane X-ray fluoroscopy. If the saddle coils are rotated around the y-axis, an open area is changed. Therefore, the saddle coils and bi-plane X-ray fluoroscopy must be rotated simultaneously. To confirm the feasibility of 3D locomotion of the microrobot, we executed a locomotion test of the microrobot in the blood vessel phantom, where the blood vessel phantom was fabricated by the rendering data from computed tomography (CT) images of the iliac artery and 3D printer.
  • Keywords
    X-ray imaging; blood vessels; coils; computerised tomography; electromagnetic actuators; medical robotics; microrobots; patient treatment; radiography; 3D printer; Helmholtz-Maxwell coils; bi-plane X-ray fluoroscopy; bi-plane X-ray imaging device; blood vessel; computed tomography images; electromagnetic actuation system; iliac artery; locomotive intravascular therapeutic microrobot; rendering data; uniform-gradient saddle coils; Blood vessels; Coils; Magnetic fields; Magnetic resonance imaging; Phantoms; Three-dimensional displays;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Biomedical Robotics and Biomechatronics (2014 5th IEEE RAS & EMBS International Conference on
  • Conference_Location
    Sao Paulo
  • ISSN
    2155-1774
  • Print_ISBN
    978-1-4799-3126-2
  • Type

    conf

  • DOI
    10.1109/BIOROB.2014.6913882
  • Filename
    6913882