DocumentCode
2605124
Title
Measure uncertainty in ultraviolet index standard
Author
Zhang, Jia-Long ; Shieh, Jen-Yu ; Wang, Zih-Yan ; Huang, Song-Yuh ; Huang, Song-Chi
Author_Institution
Dept. of Electro-Opt. Eng., Nat. Formosa Univ., Hue-Wei, Taiwan
fYear
2011
fDate
14-17 June 2011
Firstpage
552
Lastpage
555
Abstract
This paper follows International Standards Organization (ISO) by “Guide to the Expression of Uncertainty in Measurement, GUM”. In order to measure Ultraviolet Index (UVI) standard, it is necessary in calibration or test laboratory field norm on the authentication. By distance Inverse-Square Law of the Optical Radiation, We constructed measurement equation, to consider the error source when you are measuring, to evaluate the uncertainty of system. Evaluate the uncertainty of this system: The measurement distance is 75mm, ultraviolet index of the standard source is y = 11.7 UVI, effective degrees of freedom is Veff = 67, coverage factor is k = 2.0, confidence level is 95%, the standard uncertainty is uc= 0.13 UVI, expanded uncertainty is U= 0.3 UVI, the result is Y= y ± U= 11.7 ± 0.3 UVI.
Keywords
ISO standards; calibration; light sources; measurement errors; measurement uncertainty; Guide to the Expression of Uncertainty in Measurement; International Standards Organization; Inverse-Square Law of the Optical Radiation; error source; light source; measurement equation; test laboratory field norm; ultraviolet index standard; Calibration; Instruments; Light sources; Measurement uncertainty; Optical variables measurement; Temperature measurement; Uncertainty;
fLanguage
English
Publisher
ieee
Conference_Titel
Consumer Electronics (ISCE), 2011 IEEE 15th International Symposium on
Conference_Location
Singapore
ISSN
0747-668X
Print_ISBN
978-1-61284-843-3
Type
conf
DOI
10.1109/ISCE.2011.5973891
Filename
5973891
Link To Document