• DocumentCode
    2611603
  • Title

    Performance tests of a stereotactic gamma imaging system used for localization

  • Author

    Welch, Benjamin L. ; Banks, David ; St.Saviour, Torre

  • Author_Institution
    Dilon Technologies, Newport News VA 23606 USA
  • fYear
    2008
  • fDate
    19-25 Oct. 2008
  • Firstpage
    3892
  • Lastpage
    3897
  • Abstract
    The performance and accuracy of a stereotactic gamma imaging and localization system has been verified and the results of the verification are presented. The system is intended for use in localizing suspected breast cancer lesions occult to other imaging modalities. It consists of a small field-of-view gamma camera, a pair of slant-hole collimators, and a grid localization system. Tests with point sources have demonstrated that the system performs as expected and can be used to accurately determine the three dimension location of point sources. Tests with point sources have also been conducted to determine the accuracy of the slant-hole collimator system and the functionality of the software used to determine the location. The results of these tests are presented.
  • Keywords
    Breast biopsy; Breast cancer; Breast neoplasms; Cameras; Collimators; Lesions; Optical imaging; Software testing; System testing; X-ray imaging;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium Conference Record, 2008. NSS '08. IEEE
  • Conference_Location
    Dresden, Germany
  • ISSN
    1095-7863
  • Print_ISBN
    978-1-4244-2714-7
  • Electronic_ISBN
    1095-7863
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2008.4774133
  • Filename