Abstract :
We developed a prototype of TlBr semiconductor detector array for high resolution PET scanner with less than 1mm FWHM. We obtained an energy resolution of ∼20% FWHM for 511 keV and a coincidence time resolution of ∼50nsec FWHM at a detector bias 100V. This result expects the capability of application of TlBr detector array to a high resolution PET scanner. The TlBr detector arrays with the present time resolution performance can be applied to a small PET scanner for the PET study using small animals such as rats and mice.
Keywords :
Animals; Detectors; Energy resolution; Polarization; Positron emission tomography; Power engineering and energy; Prototypes; Sensor arrays; Spatial resolution; Whole-body PET;