Title :
A cone beam CT scanner without moving parts
Author :
Schomberg, Hermann ; Luhta, Randy ; Pietig, Rainer
Author_Institution :
Philips Research Europe - Hamburg, Germany
Abstract :
We propose a cone beam CT scanner without moving parts. The proposed scanner has a large number of small, stationary x-ray sources that are aligned along a specially shaped source curve. It also has a stationary detector strip with a large area and a special shaping. The sources are based on carbon nanotube field emitters. The detector is assembled from smaller modules having a flat, rectangular detector area. The special shapings of the source curve and the detector strip make it possible to collect a complete set of cone beam projections of a patient’s volume of interest in a very short time. The proposed scanner appears to be well suited for cardiac CT.
Keywords :
Assembly; Carbon nanotubes; Computed tomography; Electron beams; Europe; Optical imaging; Particle beams; Strips; X-ray detection; X-ray detectors;
Conference_Titel :
Nuclear Science Symposium Conference Record, 2008. NSS '08. IEEE
Conference_Location :
Dresden, Germany
Print_ISBN :
978-1-4244-2714-7
Electronic_ISBN :
1095-7863
DOI :
10.1109/NSSMIC.2008.4774140