Title :
The efficient realization of B mode ultrasonic Digital Scan Conversion based on radix-4 CORDIC Algorithm
Author :
Zeng, Peng ; Lu, Fanshui ; Zhao, Xingqun
Author_Institution :
Dept. of Biomed. Eng., Southeast Univ., Nanjing, China
Abstract :
Coordinate transform is the core technology in Digital Scan Conversion (DSC) of B-mode ultrasonic, and its speed mostly determines the frame rate of B-mode imaging. In the traditional coordinate transform process, ultrasonic echo data is converted from polar coordinate to rectangular directly; the process is time-consuming and the data is unsuitable for post-treatment. In this paper, we employed the CORDIC Algorithm in coordinate transform. We compared the conventional radix-2 CORDIC algorithm with the improved radix-4 CORDIC algorithm and found the later saved half iterations at the same frequency. What´s more, radix-4 CORDIC algorithm had smaller hardware consumption when using the pipelined structure. These results revealed that radix-4 CORDIC is an excellent method for high-speed DSC of B-mode ultrasonic.
Keywords :
echo; signal processing; transforms; ultrasonic applications; B mode ultrasonic digital scan conversion; B-mode imaging; DSC; coordinate transform; improved radix-4 CORDIC algorithm; ultrasonic echo data; DSC; FPGA architecture; coordinate transform; radix-4 CORDIC algorithm;
Conference_Titel :
Image and Signal Processing (CISP), 2011 4th International Congress on
Conference_Location :
Shanghai
Print_ISBN :
978-1-4244-9304-3
DOI :
10.1109/CISP.2011.6100702