• DocumentCode
    2613577
  • Title

    Adaptive upsampling with shift method for windmill artifact correction in multislice helical CT

  • Author

    Zamyatin, Alexander A.

  • Author_Institution
    Toshiba Medical Research Institute USA, Vernon Hills, IL 60061, USA
  • fYear
    2008
  • fDate
    19-25 Oct. 2008
  • Firstpage
    4296
  • Lastpage
    4298
  • Abstract
    Compared to data-domain z-filtering, the proposed “up-sampling with shift” method provides stronger artifact suppression. On the other hand, it has a potential to preserve image z-resolution by incorporating adaptive techniques. The proposed method has some similarities with zFFS techniques described in [6],[7]. The difference of our approach is that it is simple to implement and does not require hardware changes, like FFS, or changes in the backprojection module. It does not require reconstruction of several image slices for each final image, like image-domain z-filtering, so reconstruction speed is not sacrificed. In this work we apply the proposed upsampling method before reconstruction. However, for improved efficiency we suggest applying the proposed method after the convolution, just before the backprojection.
  • Keywords
    Blades; Collimators; Computed tomography; Degradation; Detectors; Image quality; Image reconstruction; Interpolation; Kernel; Nuclear and plasma sciences;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium Conference Record, 2008. NSS '08. IEEE
  • Conference_Location
    Dresden, Germany
  • ISSN
    1095-7863
  • Print_ISBN
    978-1-4244-2714-7
  • Electronic_ISBN
    1095-7863
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2008.4774233
  • Filename