Reduction of CT artifacts due to respiratory motion in a slowly rotating SPECT/CT
Author
Erlandsson, Kjell ; Nunez, Margarita ; Kruppa, Declan ; Hutton, Brian F.
Author_Institution
Institute of Nuclear Medicine, University College London, UK
fYear
2008
fDate
19-25 Oct. 2008
Firstpage
3775
Lastpage
3778
Abstract
The GE Hawkeye-4 is a low-dose CT system mounted on the gantry of a SPECT scanner. With a scanning time of 23 s per full rotation, which corresponds to 4–5 normal breathing cycles, respiratory motion may induce image artifacts in thorax studies. We have developed a method to reduce these artifacts by correcting for the axial component of the respiratory motion, utilizing respiratory phase information concerning the breathing pattern, obtained by the use of a pneumatic belt-system. The method is based on segmentation of the projection data into subsets corresponding to different motion-amplitudes, followed by generation of low quality, but motion-free image estimates. The method has been tested with simulations based on a digital phantom and also with simulations based on real CT data. Our results show that significant reduction in artifacts can be achieved by correcting for the axial component of the respiratory motion. In conclusion, this is a promising method for reduction of respiratory motion artifacts in a slowly rotating SPECT/CT system.
Keywords
Computed tomography; Image reconstruction; Image segmentation; Imaging phantoms; Manufacturing; Motion estimation; Nuclear and plasma sciences; Testing; Thorax; Timing;
fLanguage
English
Publisher
ieee
Conference_Titel
Nuclear Science Symposium Conference Record, 2008. NSS '08. IEEE
Conference_Location
Dresden, Germany
ISSN
1095-7863
Print_ISBN
978-1-4244-2714-7
Electronic_ISBN
1095-7863
Type
conf
DOI
10.1109/NSSMIC.2008.4774279
Filename
4774279
Link To Document