X-ray cone-beam breast computed tomography: Phantom studies
Author :
Russo, P. ; Lauria, A. ; Mettivier, G. ; Montesi, M.C.
Author_Institution :
Dipartimento di Scienze Fisiche, UniversitÃ\xa0 di Napoli Federico II, Italy
fYear :
2008
fDate :
19-25 Oct. 2008
Firstpage :
4803
Lastpage :
4810
Abstract :
We present the design and preliminary imaging evaluation of an X-ray cone-beam breast computed tomography (CBBCT) apparatus, assembled for laboratory test of various optimization techniques for clinical CBBCT. The prototype is characterized by the computer-control of X-ray tube (35–80 kVp, 0.25 mA), CsI:Tl flat panel detector (12×12 cm2 area, 50 μm pitch), translation and rotation stages with eight degrees of freedom, fan-beam or cone-beam FDK backprojection. First tomographic images, obtained by rotating 14-cm diameter PMMA breast phantoms are presented.
Keywords :
Assembly; Breast; Computed tomography; Design optimization; Imaging phantoms; Laboratories; Optical imaging; Prototypes; Testing; X-ray imaging;