DocumentCode :
2615588
Title :
Optimization of a monolithic detector block design for a prototype human brain PET scanner
Author :
Mendes, Pedro Rato ; Bruyndonckx, Peter ; Castro, Mario Cañadas ; Li, Zhi ; Perez, Jose Manuel ; Martìn, Iciar Sarasola
Author_Institution :
Ciemat - Centro de Investigaciones Energéticas, Medioambientales y Tecnológicas, Avda. Complutense 22, 28040 Madrid, Spain
fYear :
2008
fDate :
19-25 Oct. 2008
Firstpage :
4927
Lastpage :
4930
Abstract :
We are presently developing a novel PET scanner for human brain functional imaging based on monolithic scintillator crystals read by APD matrices, capable of being inserted into an MRI system. In this work we report on the detailed study that has been made of the design of the detector blocks, aiming at defining the most suitable geometrical and readout configuration for optimizing the overall performance of the entire scanner. Both parallel and trapezoidal geometries have been simulated, featuring two layers of active scintillator material with different or similar thickness and APD readout on the front or back side. Results of this study indicate that a trapezoidal geometry with equal thickness of both layers is the best solution for improving the expected scanner performance.
Keywords :
Computed tomography; Crystals; Design optimization; Detectors; Geometry; Humans; Magnetic resonance imaging; Positron emission tomography; Prototypes; Spatial resolution;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Nuclear Science Symposium Conference Record, 2008. NSS '08. IEEE
Conference_Location :
Dresden, Germany
ISSN :
1095-7863
Print_ISBN :
978-1-4244-2714-7
Electronic_ISBN :
1095-7863
Type :
conf
DOI :
10.1109/NSSMIC.2008.4774345