DocumentCode :
2615602
Title :
CZT gamma camera with pinhole collimator: Spectral measurements
Author :
Blevis, Ira ; Tsukerman, Leonid ; Volokh, Lana ; Hugg, James ; Jansen, Floris ; Bouhnik, Jean Paul
Author_Institution :
GE Healthcare, Haifa, Israel
fYear :
2008
fDate :
19-25 Oct. 2008
Firstpage :
4931
Lastpage :
4932
Abstract :
To date, CZT has been investigated for a variety of nuclear imaging applications using mostly parallel hole collimators. The image quality realized has evolved to make these configurations competitive with the mature conventional configurations using scintillators such as NaI. However the full potential of CZT detectors is still evolving. In our investigations we have found that there are significant additional advantages of the combination of monolithic CZT detectors with pinhole collimators over the known advantages of CZT detectors by themselves. In particular we show here the improved energy response spectrum for a selection of common nuclear medicine isotopes. The improved energy resolution is important to quality control, image fidelity and stability, and new applications such as dual isotope imaging.
Keywords :
Cameras; Collimators; Detectors; Energy resolution; Image quality; Isotopes; Nuclear imaging; Nuclear medicine; Quality control; Stability;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Nuclear Science Symposium Conference Record, 2008. NSS '08. IEEE
Conference_Location :
Dresden, Germany
ISSN :
1095-7863
Print_ISBN :
978-1-4244-2714-7
Electronic_ISBN :
1095-7863
Type :
conf
DOI :
10.1109/NSSMIC.2008.4774346
Filename :
4774346