DocumentCode
2615882
Title
Simulation of mechanical misalignments in a cone-beam micro-CT system
Author
Vidal-Migallón, I. ; Abella, M. ; Sisniega, A. ; Vaquero, J.J. ; Desco, M.
Author_Institution
Unidad de Medicina y Cirug?a Experimental, Hospital General Universitario Gregorio Mara??n, Spain
fYear
2008
fDate
19-25 Oct. 2008
Firstpage
5007
Lastpage
5009
Abstract
X-ray CT images usually show artefacts due not only to physical effects -e.g., beam hardening-, but also to misalignments that remain after mechanical calibration. These artefacts become particularly noticeable in the case of high spatial resolution systems and in hybrid systems, such as PET-CT, SPECT-CT scanners, which rely on a correct registration of emission and CT data. Hence, slight mechanical misalignments affect the quality of the CT images and any attenuation correction methods or further quantification based on them. We implemented a computer simulator of these artefacts on a cone-beam, flat-panel based micro-CT scanner. Using this simulator, we studied the effect of these different misalignments (pitch and roll tilts, skew and shifts) on reconstructed images.
Keywords
Attenuation; Calibration; Computational modeling; Computed tomography; Detectors; Image reconstruction; Imaging phantoms; Nuclear and plasma sciences; Projection algorithms; Spatial resolution;
fLanguage
English
Publisher
ieee
Conference_Titel
Nuclear Science Symposium Conference Record, 2008. NSS '08. IEEE
Conference_Location
Dresden, Germany
ISSN
1095-7863
Print_ISBN
978-1-4244-2714-7
Electronic_ISBN
1095-7863
Type
conf
DOI
10.1109/NSSMIC.2008.4774364