DocumentCode
2616458
Title
Iterative EM reconstruction of cardiac small animal PET images using system point spread function modelling and MAP with anatomical priors
Author
Spinelli, Antonello E. ; Fiacchi, Giacomo ; D´Ambrosio, Daniela ; Boschi, Stefano ; Franchi, Roberto ; Marengo, Mario
Author_Institution
Servizio di Fisica Sanitaria, Policlinico S. Orsola - Malpighi, Bologna, Italy
fYear
2008
fDate
19-25 Oct. 2008
Firstpage
5147
Lastpage
5152
Abstract
The goal of this work was to improve the image quality of small animal PET images by introducing in the reconstruction process the true system point spread function (PSF) and an anatomical image prior (MAP). The algorithm was tested on a mouse heart digital phantom (myocardium and ventricles) and on real acquired images of the Derenzo phantom and of a real mouse. For both the acquisitions two radiotracers were used, 18-FDG and 68Ga-chloride. A comparison between standard EM reconstruction and EM with PSF modelling and anatomical prior was performed. The system PSF was assumed to be a Gaussian function and its Full Width Half Maximum (FWHM) was modelled to be spatially variant in order to simulate the different spatial resolution inside the scanner field of view. A visual comparison of the images reconstructed with the standard EM and with the proposed image reconstruction method showed that the reconstructed images look much sharper and are very close to the true ones when using EM with PSF modelling and anatomical prior.
Keywords
Animals; Heart; Image quality; Image reconstruction; Imaging phantoms; Mice; Myocardium; Positron emission tomography; Spatial resolution; Testing; MAP; anatomical prior; moby; point spread function;
fLanguage
English
Publisher
ieee
Conference_Titel
Nuclear Science Symposium Conference Record, 2008. NSS '08. IEEE
Conference_Location
Dresden, Germany
ISSN
1095-7863
Print_ISBN
978-1-4244-2714-7
Electronic_ISBN
1095-7863
Type
conf
DOI
10.1109/NSSMIC.2008.4774395
Filename
4774395