Abstract
We present an efficient two-step Hilbert reconstruction method working on M-lines which allows for mathematical accurate reconstruction using all the data on a standard cylindrical detector at maximum pitch. We give an overview over the problem of exact reconstruction using redundant data, describe the algorithm in detail and present results showing a notable reduction of image noise over algorithms which do not use redundant data, without compromising resolution.
Keywords
Algorithm design and analysis; Computed tomography; Detectors; Filtering algorithms; Image reconstruction; Image resolution; Medical services; Noise reduction; Nuclear and plasma sciences; Reconstruction algorithms;