• DocumentCode
    2617865
  • Title

    Investigation of detector coverage for exact reconstruction with general circle+curve trajectory

  • Author

    Zamyatin, Alexander A. ; Katsevich, Alexander

  • Author_Institution
    Toshiba Medical Research Institute USA, Vernon, Hills, IL 60061, USA
  • fYear
    2008
  • fDate
    19-25 Oct. 2008
  • Firstpage
    5423
  • Lastpage
    5425
  • Abstract
    We show that the general circle+curve algorithm for CT reconstruction does not impose any additional requirements on the detector coverage, compared to the approximate circle-only reconstruction, provided we use a cylindrical detector. That is, we show that there is no longitudinal truncation of the filtering lines with the proposed general circle+curve algorithm. In practice it means that no extrapolation (or, alternatively, restriction of the reconstruction filed of view) is required for the exact image reconstruction.
  • Keywords
    Biomedical imaging; Computed tomography; Equations; Extrapolation; Filtering algorithms; Image reconstruction; Nuclear and plasma sciences; Object detection; X-ray detection; X-ray detectors;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium Conference Record, 2008. NSS '08. IEEE
  • Conference_Location
    Dresden, Germany
  • ISSN
    1095-7863
  • Print_ISBN
    978-1-4244-2714-7
  • Electronic_ISBN
    1095-7863
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2008.4774481
  • Filename
    4774481