Performance of Argon-Xenon mixtures in a gas proportional scintillation counter for the 0.1-10 keV X-ray region
Author :
Carmo, S. J c do ; Borges, F.I.G.M. ; Conde, C.A.N.
Author_Institution :
Departamento de FÃ\xadsica da Universidade de Coimbra, P-3004-516, Portugal
fYear :
2008
fDate :
19-25 Oct. 2008
Firstpage :
898
Lastpage :
901
Abstract :
The performance of a gas proportional scintillation counter filled with either pure Xe or Ar-Xe mixtures is described for the 0.1-10 keV X-ray energy range. It is shown that the spectra tail distortion exhibited by Xe filled gaseous detectors for soft X-rays below 2 keV is reduced if Ar-Xe mixtures are used. The peak-to-valley ratio increases by a factor of about 2 or 3 as the Xe concentration is reduced from 100% to 5%. Moreover, the energy resolutions for such mixtures are similar to the ones for pure Xe.
Keywords :
Electromagnetic wave absorption; Energy resolution; Instruments; Nuclear and plasma sciences; Radiation detectors; Scintillation counters; Solid scintillation detectors; Tail; X-ray detection; X-ray detectors; Ar-Xe mixtures; X-ray detector; gas proportional scintillation counters; gaseous radiation detectors; soft X-rays; xenon;