DocumentCode
2620437
Title
Upgrade of position sensitive detector system for small angle neutron scattering
Author
Engels, R. ; Clemens, U. ; Erven, W. ; Frielinghaus, H. ; Heiderich, M. ; Noldgen, H. ; Kemmerling, G. ; Lovenich, H.-W. ; Ossovyi, V. ; Pipich, V. ; Radulescu, A. ; Schelten, J.
Author_Institution
Forschungszentrum Jÿlich GmbH, Zentralinstitut fÿr Elektronik, D-52425, Germany
fYear
2008
fDate
19-25 Oct. 2008
Firstpage
1206
Lastpage
1208
Abstract
At the small angle neutron scattering instrument KWS2 of the new research reactor FRM II at Munich an upgrade of the detector system is currently under development. In order to improve the count rate capability, the readout electronics of the large scintillation detector is to be replaced by a modern state-of-the-art electronics, which implements a shorter pulse processing and higher bandwidth for transport of the data. An overall dead time of 1 μs is aimed at with the new design. First measurements with the new readout-electronics have been carried out, showing a considerable increase of the count rate capability.
Keywords
Collimators; Inductors; Instruments; Lenses; Neutron spin echo; Photomultipliers; Position sensitive particle detectors; Readout electronics; Scintillation counters; Solid scintillation detectors;
fLanguage
English
Publisher
ieee
Conference_Titel
Nuclear Science Symposium Conference Record, 2008. NSS '08. IEEE
Conference_Location
Dresden, Germany
ISSN
1095-7863
Print_ISBN
978-1-4244-2714-7
Electronic_ISBN
1095-7863
Type
conf
DOI
10.1109/NSSMIC.2008.4774620
Filename
4774620
Link To Document