Title : 
Radiation detectors based on silicon monolithic telescope in medical applications
         
        
            Author : 
Fazzi, Alberto ; Agosteo, Stefano ; Pola, Andrea ; Introini, Maria Vittoria ; Varoli, Vincenzo
         
        
            Author_Institution : 
Department of Energy, Politecnico di Milano, via Ponzio 34/3, I-20131, Italy
         
        
        
        
        
        
            Abstract : 
Different silicon detectors have been designed and fabricated based on the silicon monolithic telescope concept. The surface vertical diode, the ΔE stage, is only about 2 μm thick and shows very good charge collection properties. Circular anodes with a diameter of about 10 μm allow to obtain cylindrical sensitive volumes of micrometric size. This micrometric diode has been proposed as a silicon microdosimeter with the proper corrections for the chord distribution and the tissue equivalence. The lineal energy spectrum has been successfully compared with TEPC spectra and the dose-mean lineal energy with data from literature. The present lower limit is about 10 keV μm−1 fixed by the electronic noise.
         
        
            Keywords : 
Biomedical equipment; Boron; Diodes; Dosimetry; Medical services; Neutrons; Radiation detectors; Silicon radiation detectors; Spectroscopy; Telescopes;
         
        
        
        
            Conference_Titel : 
Nuclear Science Symposium Conference Record, 2008. NSS '08. IEEE
         
        
            Conference_Location : 
Dresden, Germany
         
        
        
            Print_ISBN : 
978-1-4244-2714-7
         
        
            Electronic_ISBN : 
1095-7863
         
        
        
            DOI : 
10.1109/NSSMIC.2008.4775114