DocumentCode
2629743
Title
Temperature dependence in the long-term stability of the TlBr detector
Author
Da Costa, Fábio Eduardo ; Hamada, Margarida Mizue
Author_Institution
IPEN/CNEN-SP, Travessa R, 400, Cidade Universitária, 05508-900, Sao Paulo, Brazil
fYear
2008
fDate
19-25 Oct. 2008
Firstpage
348
Lastpage
352
Abstract
In this paper, the temperature dependence in the long term stability of the TlBr detector was studied. Systematic measurements of the pulse height spectra and counting rate were carried out. The pulse height spectra showed poor hole collection, worsening with the increase of temperature. The ratemeter measurements showed an aleatory pulse formation that may be attributed to the polarization effect and air moisture in the detector environment.
Keywords
Anodes; Electrodes; Polarization; Preamplifiers; Pulse amplifiers; Pulse measurements; Radiation detectors; Stability; Temperature dependence; Temperature measurement;
fLanguage
English
Publisher
ieee
Conference_Titel
Nuclear Science Symposium Conference Record, 2008. NSS '08. IEEE
Conference_Location
Dresden, Germany
ISSN
1095-7863
Print_ISBN
978-1-4244-2714-7
Electronic_ISBN
1095-7863
Type
conf
DOI
10.1109/NSSMIC.2008.4775183
Filename
4775183
Link To Document