DocumentCode :
2629743
Title :
Temperature dependence in the long-term stability of the TlBr detector
Author :
Da Costa, Fábio Eduardo ; Hamada, Margarida Mizue
Author_Institution :
IPEN/CNEN-SP, Travessa R, 400, Cidade Universitária, 05508-900, Sao Paulo, Brazil
fYear :
2008
fDate :
19-25 Oct. 2008
Firstpage :
348
Lastpage :
352
Abstract :
In this paper, the temperature dependence in the long term stability of the TlBr detector was studied. Systematic measurements of the pulse height spectra and counting rate were carried out. The pulse height spectra showed poor hole collection, worsening with the increase of temperature. The ratemeter measurements showed an aleatory pulse formation that may be attributed to the polarization effect and air moisture in the detector environment.
Keywords :
Anodes; Electrodes; Polarization; Preamplifiers; Pulse amplifiers; Pulse measurements; Radiation detectors; Stability; Temperature dependence; Temperature measurement;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Nuclear Science Symposium Conference Record, 2008. NSS '08. IEEE
Conference_Location :
Dresden, Germany
ISSN :
1095-7863
Print_ISBN :
978-1-4244-2714-7
Electronic_ISBN :
1095-7863
Type :
conf
DOI :
10.1109/NSSMIC.2008.4775183
Filename :
4775183
Link To Document :
بازگشت