• DocumentCode
    2630010
  • Title

    A novel HV system for multi specimens aging tests under partial discharges and temperature

  • Author

    Candela, R. ; Romano, P. ; Schifani, R.

  • Author_Institution
    Dipartimento di Ingegneria Elettrica, Palermo Univ., Italy
  • Volume
    2
  • fYear
    2000
  • fDate
    2000
  • Firstpage
    732
  • Abstract
    In this paper a novel high voltage (HV) system for testing five specimens in parallel under different combinations of voltage and temperature is presented. When partial discharges (PD) are present, the system allows under a computer supervision the cyclic acquisition of PD signals coming out from each specimen by mean of a new partial discharge phase resolved analyser (PDPRA) built in our laboratory. Each specimen is provided by a suitable circuit breaker, ad hoc designed, that disconnects the high voltage (HV) side of the broken-down specimen, without the central HV supply is interested. The specimen temperature regulation (up to 100°C) is performed by a pt100 thermo regulator in a silicon oil bath. Finally, some tests have been performed to check the absence of any mutual PD interference between the specimens
  • Keywords
    ageing; high-voltage techniques; insulation testing; partial discharge measurement; 100 C; circuit breaker; computer aided instrumentation; dielectric material; electrical insulation; high voltage system; multi-specimen aging test; partial discharge phase resolved analyser; temperature regulation; Aging; Circuit breakers; Laboratories; Partial discharges; Regulators; Signal analysis; Signal resolution; System testing; Temperature; Voltage;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Electrical Insulation and Dielectric Phenomena, 2000 Annual Report Conference on
  • Conference_Location
    Victoria, BC
  • Print_ISBN
    0-7803-6413-9
  • Type

    conf

  • DOI
    10.1109/CEIDP.2000.884062
  • Filename
    884062