DocumentCode
2633225
Title
Tracer kinetic parametric imaging in PET
Author
Carson, Richard E.
Author_Institution
Nat. Inst. of Health, Bethesda, MD, USA
fYear
2004
fDate
15-18 April 2004
Firstpage
611
Abstract
Dynamic PET image acquisition is combined with tracer kinetic modeling techniques to produce quantitative estimates of physiological parameters on the regional or voxel level. Current and future challenges in this area are explored. As the resolution of PET scanners improves, correction for subject motion during the long acquisition periods must be applied. Direct measurement of an appropriate input function for the models is being replaced by image-based analyses. Development of improved, rapid parameter estimation algorithms is necessary to assure unbiased parametric images with low statistical noise. Finally, direct incorporation of the kinetic model into the reconstruction process is likely to be the ultimate path for production of tracer kinetic parametric images for PET.
Keywords
biomechanics; parameter estimation; positron emission tomography; dynamic PET image acquisition; image reconstruction; image-based analyses; low statistical noise; parameter estimation; subject motion correction; tracer kinetic parametric imaging; unbiased parametric images; Attenuation; Blood flow; Brain modeling; Image analysis; Image registration; Kinetic theory; Motion analysis; Parameter estimation; Positron emission tomography; Production;
fLanguage
English
Publisher
ieee
Conference_Titel
Biomedical Imaging: Nano to Macro, 2004. IEEE International Symposium on
Print_ISBN
0-7803-8388-5
Type
conf
DOI
10.1109/ISBI.2004.1398612
Filename
1398612
Link To Document