DocumentCode
2634783
Title
Accelerated parallel magnetic resonance imaging by adaptive k-space sampling
Author
Aggarwal, Nitin ; Bresler, Yoram
Author_Institution
Dept. of Electr. & Comput. Eng., Illinois Univ., Champaign, IL, USA
fYear
2004
fDate
15-18 April 2004
Firstpage
892
Abstract
Parallel MR imaging methods like SMASH, SENSE etc. use multiple receiver coils to accelerate the imaging process by reducing the Fourier space sampling requirement. In this paper, we propose a method for selecting the optimal k-space sampling scheme for parallel MR imaging that minimizes the expected reconstruction error given coil sensitivities and desired acceleration factor. We demonstrate that higher acceleration factors are feasible by spacing the k-space non-uniformly instead of at an integer multiple of the Nyquist spacing.
Keywords
Fourier transforms; Nyquist criterion; biomedical MRI; image reconstruction; medical image processing; Fourier space sampling; Nyquist spacing; accelerated parallel magnetic resonance imaging; acceleration factor; adaptive k-space sampling; receiver coil sensitivities; Acceleration; Coils; Data acquisition; Fourier transforms; Image reconstruction; Image resolution; Image sampling; Magnetic resonance imaging; Sampling methods; Two dimensional displays;