Calibration of a directional hydrophone in a hydroacoustic tank using the Prony method
Author :
Schlosberg, David
Author_Institution :
Hermes Electron. Inc., Dartmouth, NS, Canada
Volume :
2
fYear :
1997
fDate :
6-9 Oct 1997
Firstpage :
1339
Abstract :
We have shown that the Prony method, used with a stepped-sinusoid pulse, is suited to our directional hydrophone calibration. We are able to drop our lower frequency limit from 750 Hz (3 cycles) to 200 Hz (0.8 cycles). Further study into pole convergence methods and more recent Prony method research is necessary to find out whether the poles of the test device can be ascertained directly. Such a possibility should produce a faster calibration and more useful information for modeling purposes
Keywords :
calibration; hydrophones; 200 Hz; 750 Hz; Prony method; calibration; directional hydrophone; hydroacoustic tank; lower frequency limit; pole convergence methods; stepped-sinusoid pulse; test device; Calibration; Difference equations; Discrete Fourier transforms; Electron tubes; Electronic equipment testing; Frequency; Pressure measurement; Resonance; Sonar equipment; Steady-state;