DocumentCode
2635582
Title
High accurate Creep compensation method for Load Cell
Author
Makabe, Makoto ; Kohashi, Toru
Author_Institution
Yamato Scale Co., Ltd., Akashi
fYear
2007
fDate
17-20 Sept. 2007
Firstpage
29
Lastpage
36
Abstract
Since low capacity or stainless steel load cells are subject to larger creep errors than medium or high capacity aluminum load cells, the conventional analog compensation method cannot fully cancel out the errors. In particular, the rate of change with creep is high immediately after load application or removal. While OIML R 60 has no stipulation regarding changes in creep immediately after load application or removal, it is important and useful for automatic and non-automatic weighing scales to compensate so as to minimize the creep error during those periods of time. We have developed an accurate compensation method for creep error that occurs immediately after load application or removal on load cells.
Keywords
aluminium; creep; stainless steel; aluminum load cells; automatic weighing scales; creep error; high accurate creep compensation method; stainless steel load cells; Aluminum; Bonding; Capacitive sensors; Creep; Equations; Inorganic materials; Research and development; Steel; Timing; Weight measurement; Creep; Load cell; Start point;
fLanguage
English
Publisher
ieee
Conference_Titel
SICE, 2007 Annual Conference
Conference_Location
Takamatsu
Print_ISBN
978-4-907764-27-2
Electronic_ISBN
978-4-907764-27-2
Type
conf
DOI
10.1109/SICE.2007.4420945
Filename
4420945
Link To Document