DocumentCode
2639300
Title
Computer simulation of eddy current loss reduction for rechargeable cardiac pacemaker
Author
Sato, T. ; Sato, F. ; Matsuki, H. ; Sato, Takao
Author_Institution
Dept. of Electr. & Commun. Eng., Tohoku Univ., Japan
fYear
2005
fDate
12-15 May 2005
Firstpage
120
Lastpage
121
Abstract
We have studied about an electric power transmission system for implanted medical device such as cardiac pacemaker. Inside the body is heated by eddy current loss of the titanium case when power transmitted. However the eddy current loss always exists when power transmitted Purpose of this paper is to discuss about eddy current reduction by changing a shape of the titanium case, based on a computer simulation. The eddy current toss decreased substantially to 40 % when the titanium case was divided into 2 parts. The transmission efficiency was achieved 36-52 % at coil gap of 10-20 mm.
Keywords
cardiovascular system; digital simulation; eddy current losses; pacemakers; power transmission; titanium; 10 to 20 mm; computer simulation; eddy current loss reduction; electric power transmission system; implanted medical device; rechargeable cardiac pacemaker; titanium case; Batteries; Coils; Computational modeling; Computer simulation; Eddy currents; Magnetic flux density; Pacemakers; Propagation losses; Shape; Titanium; eddy current loss; pacemaker; power transmission;
fLanguage
English
Publisher
ieee
Conference_Titel
Microtechnology in Medicine and Biology, 2005. 3rd IEEE/EMBS Special Topic Conference on
Print_ISBN
0-7803-8711-2
Type
conf
DOI
10.1109/MMB.2005.1548400
Filename
1548400