Abstract
This paper describes the development of a spectral radiance calibration standard in the shortwave infrared (SWIR) range from 1000 to 2500 nm. Gold-coated and Zenith integrating spheres were developed for the working standard. Calibration of the spectral radiance of the integrating sphere was also investigated. An independent radiometer cross-checked calibration of the spectral radiance of the integrating sphere with a consistency of 1.5%.
Keywords
calibration; infrared detectors; radiometers; SWIR range; Zenith integrating spheres; gold-coated integrating spheres; independent radiometer cross-checked calibration; integrating sphere; shortwave infrared range; spectral radiance calibration standard; Apertures; Barium; Calibration; Gold; Infrared spectra; Lamps; Optical sensors; Radiometry; Standards development; Uncertainty; SWIR; Zenith; calibration; gold-coated; integrating sphere; spectral radiance;