DocumentCode
2657692
Title
A novel structure for CdZnTe capacitive frisch grid radiation detector
Author
Yongbiao, Qian ; Wenbin, Sang ; Jun, Xia ; Jiahua, Min ; Jianyong, Teng ; Xiaoyan, Liang
Author_Institution
Sch. of Mater. Sci. & Eng., Shanghai Univ.
fYear
2006
fDate
27-28 June 2006
Firstpage
304
Lastpage
307
Abstract
CdZnTe gamma spectrometers with capacitive Frisch grid structure can potentially achieve a higher detecting resolution and their geometrical structures have an important role in optimizing the detector performance. In the paper, three geometrical structures of capacitive Frisch grid detectors were investigated by using Finite element method based on the 3-dimensional weighting potential and field distribution. Simulation results indicated the trapezoid prism model could get better detector performance, whose novel structure has not only a better weighting potential distribution but also a bigger detector volume than the cuboid models
Keywords
cadmium compounds; finite element analysis; gamma-ray spectrometers; optical prisms; particle detectors; zinc compounds; 3D field distribution; 3D weighting potential; CdZnTe; capacitive Frisch grid; finite element method; gamma spectrometers; radiation detector; trapezoid prism model; Anodes; Charge measurement; Current measurement; Dielectrics; Electrodes; Electrons; Gamma ray detection; Gamma ray detectors; Radiation detectors; Solid modeling;
fLanguage
English
Publisher
ieee
Conference_Titel
High Density Microsystem Design and Packaging and Component Failure Analysis, 2006. HDP'06. Conference on
Conference_Location
Shanghai
Print_ISBN
1-4244-0488-6
Type
conf
DOI
10.1109/HDP.2006.1707611
Filename
1707611
Link To Document