  • Abstract
    Accelerator-based neutron sources are an attractive alternative to nuclear reactors for providing epithermal neutron beams for boron neutron capture therapy. Based on clinical requirements and neutronics modeling the use of proton and deuteron induced reactions in 7Li and 9Be targets has been compared. Excellent epithermal neutron beams can be produced via the 7Li(p,n)7Be reaction at proton energies of ~2.5 MeV. An electrostatic quadrupole accelerator and an Li target, which can deliver and handle 2.5 MeV protons at beam currents up to 50 mA, are under development for an accelerator-based BNCT facility at the Lawrence Berkeley National Laboratory
  • Keywords
    electrostatic accelerators; nuclear bombardment targets; proton accelerators; radiation therapy; 2.5 MeV; 7Li target; 9Be target; BNCT; Be; Li; accelerator; beam currents; boron neutron capture therapy; electrostatic quadrupole accelerator; epithermal neutron beams; neutron sources; Acceleration; Electrostatics; Inductors; Laboratories; Lithium; Neoplasms; Neutron capture therapy; Particle beams; Proton accelerators; Tumors;