Thermal stabilized front-end PCB with active cold calibration load for L-band radiometer
Author :
Kemppainen, Sami ; Lemmetyinen, Juha ; Auer, Tuomo ; Colliander, Andreas ; Aalto, Aleksi ; Rautiainen, Kimmo ; Hallikainen, Martti
Author_Institution :
Helsinki Univ. of Technol., Helsinki
fYear :
2007
fDate :
23-28 July 2007
Firstpage :
4433
Lastpage :
4436
Abstract :
In this paper, the thermally stabilized front-end of an L-band total power receiver is presented. Applications on the L-band have become one of the most important focus points in the field of passive microwave remote sensing. Measuring environmental parameters such as ocean salinity and soil moisture remain a challenge, posing strict requirements to instrument performance. For traditional radiometers, especially the stability of front-end components is critical. The principle of the stabilization technique and the actual design are described. The design also features an on-board active cold load for calibration purposes. The presented technique aims to stabilize the front-end section PCB with heaters to sub 0.05 C level.
Keywords :
moisture; oceanography; remote sensing; soil; L-band radiometer; L-band total power receiver; cold calibration load; environmental parameters; front-end components stability; ocean salinity; passive microwave remote sensing; soil moisture; thermal stabilized front-end PCB; Calibration; L-band; Microwave radiometry; Moisture measurement; Ocean salinity; Passive microwave remote sensing; Sea measurements; Soil measurements; Soil moisture; Thermal loading; front-end; radiometer; thermal stabilization;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Geoscience and Remote Sensing Symposium, 2007. IGARSS 2007. IEEE International
Conference_Location :
Barcelona
Print_ISBN :
978-1-4244-1211-2
Electronic_ISBN :
978-1-4244-1212-9
Type :
conf
DOI :
10.1109/IGARSS.2007.4423838
Filename :
4423838
Link To Document :
بازگشت