Abstract :
The Clouds and the Earth´s Radiant Energy System (CERES) instruments for the NASA Earth Observing System each contain three broadband scanning radiometers operating in bands at wavelengths from 0.3 to 5 μm, 8 to 12 μm and 0.3 to >150 μm. The primary radiance standard for CERES ground calibration is an absolute blackbody. This standard provides radiance calibration in the long wave spectral region over a blackbody temperature range of 200 K to 320 K. A cryogenic active cavity radiometer operating at liquid helium temperature is used to transfer the radiance scale of the blackbody to a secondary, shortwave calibration source. The longwave to shortwave calibration transfer makes use of the spectrally flat absorptance of the ACR across the entire spectral region. The thermal properties of the copper receiver cavity at 4 K permit a noise equivalent power performance adequate for calibration of the shortwave reference source in a series of narrow (10% wide) spectral bands which span the shortwave spectral region. The narrow spectral regions greatly reduce the errors in the relative spectral response knowledge of the shortwave source. Results are presented for a transfer that was made during validation tests of the ground calibration system in preparation for the CERES Functional Test Model instrument calibration
Keywords :
artificial satellites; calibration; geophysical equipment; geophysical techniques; infrared imaging; radiometry; submillimetre wave measurement; 0.3 to 150 mum; CERES; Clouds and the Earth´s Radiant Energy System; EHF radiometry; Functional Test Model instrument calibration; UV ultraviolet; blackbody standard; calibration; cryogenic active cavity radiometer; geophysical measurement technique; land surface terrain mapping; optical imaging; satellite method; shortwave reference standard; sub mm; submillimetre; visible infrared IR radiometry; Calibration; Clouds; Cryogenics; Earth Observing System; Instruments; Land surface temperature; NASA; Radiometers; System testing; Temperature distribution;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Geoscience and Remote Sensing Symposium, 1994. IGARSS '94. Surface and Atmospheric Remote Sensing: Technologies, Data Analysis and Interpretation., International
Conference_Location :
Pasadena, CA
Print_ISBN :
0-7803-1497-2
Type :
conf
DOI :
10.1109/IGARSS.1994.399719
Filename :
399719
Link To Document :
بازگشت