Title :
A comparison of measured and calculated FTIR spectral radiance
Author :
Han, Yong ; Churnside, James H. ; Shaw, Joseph A. ; Westwater, Ed R.
Author_Institution :
Environ. Technol. Lab., NOAA, Boulder, CO, USA
Abstract :
Data from the ETL ground-based Fourier Transform Infrared Radiometer (FTIR) were taken during the FIRE II (Coffeyville, Kansas, 1991) and PROBE (Kavieng, Papua New Guinea, 1993) experiments. In this paper calibration problems encountered during the two experiments are discussed. The FTIR spectral measurements are compared with model (FASCODE) predictions. The model agrees with the measurements with a maximum uncertainty of 8 K of the equivalent brightness temperature
Keywords :
Fourier transform spectroscopy; atmospheric humidity; atmospheric techniques; atmospheric temperature; calibration; radiometry; remote sensing; Coffeyville Kansas; FASCODE; FIRE II; FTIR; Fourier Transform Infrared Radiometer; IR radiometry; Kavieng; PROBE; Papua New Guinea; United States; atmosphere humidity; calibration; equivalent brightness temperature; measurement technique; meteorology; remote sensing; spectral radiance; Atmospheric measurements; Atmospheric modeling; Calibration; Fires; Fourier transforms; Instruments; Nitrogen; Probes; Radiometry; Remote sensing;
Conference_Titel :
Geoscience and Remote Sensing Symposium, 1994. IGARSS '94. Surface and Atmospheric Remote Sensing: Technologies, Data Analysis and Interpretation., International
Conference_Location :
Pasadena, CA
Print_ISBN :
0-7803-1497-2
DOI :
10.1109/IGARSS.1994.399723