DocumentCode
2687889
Title
Calibrating a Six-Port Reflectometer Using Three Known Loads
Author
Bialkowski, M.E. ; Woods, G.S.
fYear
1985
fDate
4-6 June 1985
Firstpage
589
Lastpage
592
Abstract
A technique for calibrating an arbitrary six-port network using nine unknown loads and three known loads is described. The technique is similar to the self-calibration technique first developed by Engen. However, it uses full six-port to four-port reduction so that it is quite insensitive to errors, The theory is illustrated with typical measurement results showing an accuracy of better than 1.0% achievable without optimization of the six-port to four-port reduction constants.
Keywords
Australia; Calibration; Circuits; Detectors; Equations; Frequency; Performance evaluation; Phase measurement; Power measurement; Reflection;