• DocumentCode
    2691086
  • Title

    Calibrating a Dual Six-Port or Four-Port for Measuring Two-Ports with Any Connectors

  • Author

    Hoer, C.A. ; Engen, G.F.

  • fYear
    1986
  • fDate
    2-4 June 1986
  • Firstpage
    665
  • Lastpage
    668
  • Abstract
    A technique is described for calibrating a dual six-port & four-port ANA so that the scattering parameters of two-port devices having any combination of connectors can be measured. The technique is a generalization of the "thru-reflect-line" (TRL) calibration technique in which the "thru" is replaced with a second length of precision transmission line.
  • Keywords
    Calibration; Connectors; Detectors; NIST; Power measurement; Power transmission lines; Reflection; Scattering parameters; Testing; Transmission line measurements;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Microwave Symposium Digest, 1986 IEEE MTT-S International
  • Conference_Location
    Baltimore, MD, USA
  • ISSN
    0149-645X
  • Type

    conf

  • DOI
    10.1109/MWSYM.1986.1132276
  • Filename
    1132276