  • Abstract
    A pneumatically actuated tunable liquid micro-lens is developed for integration in an endoscopic optical coherence tomography (OCT) system allowing dynamic focusing along the scan depth. An evaluation of the optical performance, particularly transverse resolution, is presented and its utility in OCT discussed. A transverse resolution of about 15 mum which does not vary with a scan depth of 9 mm scan depth has been obtained, performance which cannot be obtained with a fixed focal length lens.
  • Keywords
    bioMEMS; biomedical optical imaging; endoscopes; image resolution; membranes; microlenses; optical design techniques; optical focusing; optical tomography; optical tuning; polymer films; OCT-probe design; biological imaging; biomedical optics; dynamic focusing; fixed focal length lens; high lateral resolution endoscopic OCT system; optical coherence tomography; scan depth; transverse resolution; tunable multimicrolens system; Apertures; Biomembranes; High-resolution imaging; Image resolution; Integrated optics; Lenses; Optical imaging; Optical microscopy; Surface fitting; Tomography; Endoscopic optical coherence tomography; micro-lens; optical optimization; polydimethylsiloxane (PDMS);
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Optical MEMs and Nanophotonics, 2008 IEEE/LEOS Internationall Conference on
  • Conference_Location
    Freiburg
  • Print_ISBN
    978-1-4244-1917-3
  • Electronic_ISBN
    978-1-4244-1918-0
  • Type

    conf

  • DOI
    10.1109/OMEMS.2008.4607820
  • Filename
    4607820